ClearanceBrief
Recall Z-0121-06 ยท IZL System, X-Ray, Mobile

Siemens Medical Solutions USA, Inc

Siremobil Compact L Mobile X-Ray System

DateNov 2, 2005 (initiated Aug 1, 2005)
StatusTerminated
ReasonScrews may become loose and fall into the OR field during surgery.
Root causeOther
ActionThe recalling firm has their Service Engineers visiting each site to check the screws on the window and secure them with Loctite per Update Instructions SP022/05/S. The Service Engineer is also informing the account of the problem during the visits.
Quantity35 units
DistributionThe products were shipped nationwide to a distributor, a physician, and medical facilities
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963093SIREMOBIL COMPACT
Siemens Medical Solutions USA, Inc., Sep 3, 1996