Recall Z-0413-06 ยท KPS System, Tomography, Computed, Emission

GE Healthcare

GE Millennium Systems: MC, MyoSight. Dual Head Gamma Camera. A compact cardiac-optimized nuclear imaging system. It is comprised of 2 digital CSE (Correlated Signal Enhancement) detectors in variable geometries mounted on a pass-through ring gantry, a compact imaging table, and a GENIE Acquisition station.

DateJan 24, 2006 (initiated Apr 18, 2005)
StatusTerminated
ReasonIn certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause the locking mechanism to unlock itself.
Root causeOther
ActionCustomer letter is being sent to all customers who have an affected system. The letter alerts customers of the potential release of the locking mechanism and requires a Weekly Collimator Exchange Procedure to avoid release of the locking latch until the permanent solution is made available.
Quantity1421 total for all affected Millenium Systems
Distributionnationwide including Puerto Rico, and to countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Denmark, Egypt, France, Germany, Iceland, India, Israel, Italy, Jordan, Korea, Kuwait, Luxembourg, Martinique, Mexico, New Zealand, Philippines Romania, Saudi Arabia, Sloveni