Recall Z-1592-05 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Epidural Infusion Pumps sold under the following labels; a) Baxter 6060E Epidural Infusion Pump, product code 2L9351, 2L9351P and 2L9351R; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore b) Sabratek 6060 Epidural Pain Management Pump, product codes 606000-40E and 606000-40EL; Manufactured by Sabratek Corp., Skokie, IL 60076, which was purchased b

DateSep 28, 2005 (initiated Aug 17, 2005)
StatusTerminated
ReasonPotential overinfusion in the Auto-Ramp mode with software versions 3.00 or higher
Root causeOther
ActionBaxter sent the 8/17/05 Urgent Product Recall letter to all Baxter 6060E Infusion Pump customers to inform them of a software anomaly present in software versions 3.00 and higher, which allows the pump to overinfuse in the Auto-Ramp mode. The end users were also informed that these epidural pumps, identified by the yellow housing, can no longer be supported due to a lack of availability of spare parts and Baxter is unable to replace the software with an alternate version. As a result, Baxter requested the accounts to remove the pumps from service and return them to Baxter. The accounts were
DistributionNationwide including Puerto Rico, and intenationally to Canada, China, Belgium, Hong Kong and Saudi Arabia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0026796060 MULTI-THERAPY PUMP, MODEL 2M9832; 6060 EPIDURAL PUMP, MODEL 2L9351
Baxter Healthcare Corp, Sep 18, 2000
K941984SABRATEK 6060 HOMERUN VOLUMETRIC INFUSION PUMP
Sabratek Corp., Jul 19, 1994