Recall Z-0022-06 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems North America Co. Phillips

Interis Allura BiPlane X-Ray System

DateAug 31, 2005 (initiated Jul 27, 2005)
StatusTerminated
Reason1) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release of the joystick.
Root causeOther
ActionA letter issued on 7/25/05 advising customers of problem. The letter advised that an upgrade will be performed.
Quantity43 units
DistributionDevices were distributed to hospitals/medical centers nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002016PHILIPS INTEGRIS ALLURA
Philips Medical Systems North America, Inc., Sep 6, 2000