ClearanceBrief
Recall Z-1437-05 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

Baxter Interlink System Huber Needle Extension Set, product code 2N3712; An Rx sterile, nonpyrogenic fluid pathway with Interlink Y-Injection Site, Needle Gauge 20, Needle Length 1'', Total Volume 0.75 mL, Total Length 10''; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.

DateAug 30, 2005 (initiated Aug 3, 2005)
StatusTerminated
ReasonThe extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
Root causeOther
ActionBaxter sent Urgent Product Recall letters dated 8/3/05 to their customers, to the attention of the Director of Materials Management, via first class mail on the same date. The accounts were informed of the microbore female luers that may crack during use, resulting in the possible under delivery or incomplete delivery of medication. The accounts were instructed to examine their inventory for the affected lots of product, remove it from inventory and return it to Baxter for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837. Baxter extended the recall to includ
Quantity5,326 sets
DistributionNationwide, including Puerto Rico and American Samoa, and internationally to Canada, Australia, Japan, South Korea, Saudi Arabia, India, Mexico, Venezuela, Malaysia, East Timor and New Zealand.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961495HUBER NEEDLE EXTENSION SET WITH STANDARD Y-INJECTION SITE, INTERLINK Y-INJECTION SITE, INTERLINK T-INJECTION SITE
Baxter Healthcare Corp, Jun 26, 1996