Recall Z-1440-05 ยท JQP Calculator/Data Processing Module, For Clinical Use

Beckman Coulter Inc

Aqueduct Hematology Workflow Manager, software

DateAug 10, 2005
StatusTerminated
ReasonPotentail for misidentification of historical patient results within the Aqueduct Hematology Workflow Manager software. In instances where the workflow manager software receives the same patient identifier for more than one patient the Aqueduct Workflow Manager software merges all results into one Patient File. The may cause test results to be associated with the wrong patient.
Root causeOther
ActionA Product corrective action letter to be mailed the week of August 8, 2005, to all customers directing that each patient is identified with a truly unique patient identifier.
Quantity36
DistributionAK, AZ, CA, DE, FL, IN, KS, MA, MI, MN, MO, NC, NY, OH, PA, TX, VA, WA