ClearanceBrief
Recall Z-1585-05 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Bard Peripheral Vascular Inc

Bard Luminexx 3 Biliary Stent and Delivery System

DateSep 24, 2005 (initiated Jul 21, 2005)
StatusTerminated
ReasonAngiomed Material Review Board reviewing the manufacturing process observed an increase in the internal scrap rate (failure mode: broken stent struts) of nitinol stents and determined the root cause to be a manufacturing change from polymer to aluminum fixtures.
Root causeOther
ActionRecall notification letter dated July 21, 2005, was sent to Hospital and Physician consignees.
Quantity102
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033497BARD LUMINEXX 3 BILIARY STENT AND DELIVERY SYSTEM
C.R. Bard, Inc., Feb 4, 2004