Recall Z-1598-05 ยท MRM Defibrillator, Implantable, Dual-Chamber
ela Medical Llc
Alto MSP (model 617) and Alto 2 MSP (model 627) Implantable cardioverter Defibrillator. Ventricular antitachycardia pacing. Dual chamber arrhythmia detection. biventricular pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring atrial and vent
| Date | Sep 24, 2005 (initiated Jul 19, 2005) |
| Status | Terminated |
| Reason | Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function. |
| Root cause | Other |
| Action | A Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician. |
| Quantity | 20 total devices (3-model 617 & 17-model 627) |
| Distribution | Nationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI. |