Recall Z-0517-06 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Spacelabs Medical Incorporated

Ultraview SL Capnography Module, Model 91517 A mainstream and sidestream gas analyzer designed to measure the concentration of carbon dioxide in a gas mixture, to aid in determining the patient''s ventilatory, circulatory, and metabolic status.

DateFeb 16, 2006 (initiated Jul 20, 2005)
StatusTerminated
ReasonThe transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher altitudes.
Root causeOther
ActionOn 7/20/05 Spacelabs called the consignees that received the units to notify them of the hazard and also the schedule a service call to repair the affected capnography modules. The firm will confirm that all consignees have been updated with the appropriate PCBA and will audit the device recall records for completion.
Quantity11 units
Distribution8 units shipped to France, 1 unit to Brussels, Belgium, and 2 units to a medical center in WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031124POLARIS 95000-A BEDSIDE/TRANSPORT MONITOR
Datex-Ohmeda, Inc., Dec 4, 2003