Recall Z-1423-05 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Becton Dickinson & Company

BD Vacutainer Blood Transfer Device. It consists of a Luer attached via a hub to a needle covered with a sleeve.

DateAug 30, 2005 (initiated Jul 11, 2005)
StatusTerminated
ReasonCustomer complaints state the needle had separated from the hub, resulting in blood exposure and/or dislocation of the needle.
Root causeOther
ActionA customer communication was sent by first class mail on 7/26/05, advising all consignees of the problem. The letter asks each customer to immediately examine their inventory for affected lot numbers, return product from affected lots using enclosed label and packing slip, and to notify all of their affected customers to alert them to the situation.
Quantity641200 eaches
DistributionWorldwide
FDA recordOpen on fda.gov