Recall Z-1460-05 ยท LCD Enzyme Immunoassay, Gentamicin

Ortho-Clinical Diagnostics

VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests.

DateSep 1, 2005 (initiated Jul 19, 2005)
StatusTerminated
ReasonNegatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.
Root causeOther
ActionLetters dated 7/19/2005 with instructions to process QC fluid according to the IFU when processing patient samples. When replacement product arrives, instructions are to destroy the remaining inventory of the recalled lots.
Quantity443 boxes
DistributionThe product was shipped nationwide and to Ortho-Clinical/J&J affiliates in Australia, Canada, Singapore, England, France, Germany, Italy and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042479VITROS CHEMISTRY PRODUCTS GENT, TOBRA & VANC REAGENTS; VITROS CHEMISTYR PRODUCTS CALIBRATOR KITS 11, 13 & 14; VITROS CHE
Ortho-Clinical Diagnostics, Inc., Jan 6, 2005