Recall Z-1397-05 ยท NCX System, Hypothermia, Intravenous, Cooling

Alsius Corporation

Alsius CoolGard 3000 Temperature Regulation System

DateAug 24, 2005 (initiated Jul 25, 2005)
StatusTerminated
ReasonFirm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.
Root causeOther
ActionFirm will be sending letters of caution to all users that reemphasizes labeling beginning 7/25/2005. New software is being developed which will include a change to maintain STANDBY coolant temperature regulation indefinitely, removing any risk of failure.
Quantity207
DistributionNationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malasia, Italy, Ireland, greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K014241COOLGARD 3000, COOLINE, START-UP, MODELS 3000, CL-2085B, CL-2295A, CG-500D