Recall Z-1422-05 ยท JPA System, Multipurpose For In Vitro Coagulation Studies

Diagnostica Stago, Inc.

STA-R blood coagulation analyzer. Catalog Number 57160.

DateAug 30, 2005 (initiated Jun 20, 2005)
StatusTerminated
ReasonConcerning the STA-R blood coagulation analyzer, there is a possible failure mode associated with patient samples receiving prolonged exposure time to reagents during the cuvette roll change process.
Root causeOther
ActionAlert on the problem was sent out on 6/20/2005 to all customers alerting the problem and instructions on how to prevent the problem. The units are not be to returned.
Quantity215 units
DistributionUnits were distributed to clinics and hospitals nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.