Recall Z-1587-05 ยท MRM Defibrillator, Implantable, Dual-Chamber

ela Medical Llc

Alto DR (model 614) and Alto 2 DR (model 624) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Dual chamber arrhythmia detection. Antibradycardia pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring atrial and ventri

DateSep 24, 2005 (initiated Jul 19, 2005)
StatusTerminated
ReasonDevices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
Root causeOther
ActionA Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician.
Quantity392 total devices (23-model 614 & 369-model 624)
DistributionNationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI.