ClearanceBrief
Recall Z-1038-05 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Guidant Corporation

INTELIS II (model nos. 1349, 1384, 1385, 1483, 1484, 1485 and 1499) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . INTELIS II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering both adaptive-rate therapy and conventional pacing therapy. These pacemakers provide various levels of therapeutic a

DateJul 28, 2005 (initiated Jul 18, 2005)
StatusTerminated
ReasonA hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
Root causeOther
ActionINITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed
QuantityINITIAL RECALL 228. EXPANDED RECALL 6.
Distributionworldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.