Recall Z-1105-05 ยท LZG Pump, Infusion, Insulin

Roche Diagnostics Corp.

Disetronic D-Tronplus Insulin Pump adapters; Ref 3000803; catalog reference number 0457243001 sold in packages of ten adapters.

DateAug 6, 2005 (initiated Jul 15, 2005)
StatusTerminated
ReasonThe adapter valve may not operate correctly, which may result in overdelivery of insulin in the first dose after the adapter has been changed.
Root causeOther
ActionThe firm began notifying family members of pediatric patients via phone on 7/15/05 and immediately began shipping a recall letter and 10 packs of replacement adapters to each pediatric user. Attending physicians were for these pediatric patients were notified the week of 7/18/05 via phone and follow-up letters were sent to these physicians. The firm is issuing a press release on 7/28/05. Only these replacement adapters are being sold by Disetronic to users of any age ordering new adapters.
Quantity107 pediatric patients use the device.
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043000MODIFICATION TO D-TRONPLUS
Disetronic Medical Systems AG, Dec 1, 2004