Recall Z-1380-05 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Renal Div

Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration system; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085, manufactured in Germany

DateAug 18, 2005 (initiated Jul 13, 2005)
StatusTerminated
ReasonIn the event therapy is terminated, as part of routine instrument programming or therapy maintenance, the Accura instrument may experience difficulty in performing a successful self-test (failure) upon an immediate attempt to restart therapy
Root causeDevice Design
ActionBaxter sent Important Product Information letters dated 7/13/05 to all Accura device customers, to the attention of the Director of Nursing, via first class mail. The letter informed the accounts of the on-going investigation into the potential self-test failure associated with the Accura instrument, and provided the customers with steps to follow if they encounter failed instrument self-tests. These guidance steps include: 1) Place the instrument into 'stand-by' mode. Generally, 15 minutes in 'stand-by' mode is sufficient to allow successful instrument self-test procedures. Enter 'stand-
Quantity264 units
DistributionNationwide and internationally to China and Belgium