Recall Z-1576-05 ยท MZO Assay, Enzyme Linked Immunosorbent, Hepatitis C Virus

Diasorin Inc.

ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System.

DateOct 1, 2005 (initiated Mar 9, 2004)
StatusTerminated
ReasonAn error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek ELx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay.
Root causeOther
ActionDear Customer letters, dated 03/08/2004 were sent to customers informing them of the specification error and instructing users to validate their assays manually.
Quantity128
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI & WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.