Recall Z-0171-06 ยท LCP Assay, Glycosylated Hemoglobin

Provalis Diagnostics

Product is in2it A1C In-Office Analyzer Test Cartridge. Catalogue number is 902074P (Test Cartridges) which may then be packaged as a consumables pack (902079D) or a starter pack (901077N). Product is labeled in part: ''in2it In-Office analyzer A1C Test Cartridges***For the rapid quantification of hemoglobin A1c in whole blood***Provalis Diagnostics***.''

DateNov 18, 2005 (initiated May 20, 2005)
StatusTerminated
ReasonProvalis Diagnostics discovered, following its internal stability testing program (mid-point batch testing), that occasional low values were being found across multiple lots of in2it HbA1C Test Cartridges.
Root causeOther
ActionAll consignees were initially notified by telephone with followup by FAX or email. Users/Distributors may also be visited by Provalis Sales Staff.
Quantity838 Kits of 10 test devices each
DistributionDistributed to 7 Consignees in the US one each in FL, KY, IL, LA, MI, AZ, GA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041635G5 I HBA1C TEST, G5 II HBA1C TEST
Provalis Diagnostics , Ltd., Aug 16, 2004