Recall Z-0908-05 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Guidant Corporation
CONTAK RENEWAL 2 (model H155) Heart Failure Device. Sterilized with gaseous ethylene oxide. Guidant corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.
| Date | Jun 30, 2005 (initiated Jun 17, 2005) |
| Status | Terminated |
| Reason | Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry. |
| Root cause | Other |
| Action | Dear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05. |
| Quantity | 16,000 (Approximately 11,900 remain implanted worldwide) - both products combined |
| Distribution | worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, |
| FDA record | Open on fda.gov |