Recall Z-1080-05 ยท IZI System, X-Ray, Angiographic

Argon Medical Devices, Inc

Labeling is as follows: Product Labeling: ''***Argon Medical Devices***12cc Control Syringe***FR/TR/RM w/ Reservoir***SINGLE USE ONLY***REF/REORDER NO. 093209***This device is designed for the administraion and removal of fluids.***Quantity: 250 ea.***NON-STERILE*FURTHER PROCESSING REQUIRED***Eto (ethylene oxide) Sterilization Recommended******1445 Flat Creek Road Athens, TX 75751 800-927-4669**

DateAug 3, 2005 (initiated Jun 27, 2005)
StatusTerminated
ReasonThe firm is recalling the product due to the potential for air aspiration or fluid leakage from a faulty syringe tip.
Root causeOther
ActionThe firm plans to distribute recall letters to all customers indicated on the customer list. TELEPHONE 6/27/2005. All products will be returned to Argon and disposed.
Quantity2752 syringes
DistributionTN, IL, WI, MA, VA, IN, LA, TX, CA, JAPAN, GERMANY, THE NETHERLANDS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.