ClearanceBrief
Recall Z-0970-05 ยท KGC Tube, Gastro-Enterostomy

Boston Scientific Corp

Boston Scientific brand One Step Button (percutaneous endoscopic gastrostomy tubes) Kit, 24/2.4 pull; catalog number 6851.

DateJul 7, 2005 (initiated Jun 2, 2005)
StatusTerminated
ReasonTwo blades may be heat staked onto the stiletto instead of the expected one blade.
Root causeOther
ActionCustomers were notified by a recall letter dated 6/7/05 which instructed them to return any of the recalled products on hand.
DistributionNationwide and Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K910584ONE STEP BUTTON(TM)
Applied Medical Technologies, May 23, 1991