Recall Z-0956-05 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)

Boston Scientific CRM Corp

CONTACT RENEWAL 3 AVT model numbers M150 and M155; CONTACT RENEWAL 3 AVT HE model numbers M157 and M159; CONTACT RENEWAL 4 AVT model numbers M170 and M175; CONTACT RENEWAL 4 AVT HE model numbers M177 and M179. Cardiac Resynchronization Therapy Defibrillator. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 5511

DateJun 30, 2005 (initiated Jun 17, 2005)
StatusTerminated
ReasonThe atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy.
Root causeOther
ActionFirm issued a Dr. Doctor letter dated 06/17/05. Letter informs physicians of the issue, clinical implications, recommendations and action taken by the firm.
Quantity2,204 devices (786 remain implanted worldwide)
Distributionworldwide. To the US and OUS countries to include: Argentina, Australia, Austria, Belgium, Canada, Chile, Cyprus, Denmark, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Slovenia, Spain,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.