ClearanceBrief
Recall Z-0982-05 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Toshiba American Med Sys Inc

Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002

DateJul 13, 2005 (initiated May 23, 2005)
StatusTerminated
ReasonIf the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values.
Root causeOther
ActionFirm issued a recall letter which provided a procedural workaround, a software patch will be installed during routine service visits.
Quantity106
DistributionNationwide. This is an imported product.
FDA recordOpen on fda.gov

510(k) clearances named in this recall