Recall Z-0907-05 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Guidant Corporation

CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798. USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.

DateJun 30, 2005 (initiated Jun 17, 2005)
StatusTerminated
ReasonLaboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
Root causeOther
ActionDear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05.
Quantity16,000 (Approximately 11,900 remain implanted worldwide)
Distributionworldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.