Recall Z-0980-05 ยท NBW System, Test, Blood Glucose, Over The Counter

Lifescan Inc

LifeScan brand One Touch UltraSmart Meter Owners Booklet AW 060-788-01A; All One Touch UltraSmart Meters shipped with OneTouch UltraSmart Owner''s Booklet AW 060-788-01A are affected by this Field Correction.

DateJul 13, 2005 (initiated Jun 9, 2005)
StatusTerminated
ReasonMeter labeling (Owner's Booklet and Quick Reference Guide) is incorrect as it indicates that the meter displays the message 'Do you need a snack' for glucose values between 20 and 70 mg/dL; whereas the OneTouch UltraSmart meter displays the message, 'Do you need a snack?' for glucose values from 20 through 59 mg/dL.
Root causeOther
ActionOn 6/9/05, the firm initiated the recall and their notification was via letters informing its customers of the corrective action.
Quantity584,291 units
DistributionProduct was released for distribution to 40,763 consignees worldwide. The firm provided the following distribution information: 124 wholesaler dealers, 224 distributors, 140 retailers and 40,275 consumers/users involved in the recall. The recall product was distributed in all States within the U.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021819ONE TOUCH ULTRASMART BLOOD GLUCOSE MONITORING SYSTEM
Lifescan, Inc., Aug 23, 2002