Recall Z-1338-05 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Depuy Orthopaedics, Inc.

Depuy P.F.C. Modular Knee System, curved tibial insert (CVD); polyethylene tibial insert, Sz 4, 12.5 mm; Ref 98-4154.

DateAug 12, 2005 (initiated May 26, 2005)
StatusTerminated
ReasonAn incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root causeOther
Quantity5
DistributionNationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K884796P.F.C.(TM) MODULAR KNEE SYSTEM
Johnson & Johnson Professionals, Inc., Mar 29, 1989