ClearanceBrief
Recall Z-1287-05 ยท NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing

Depuy Orthopaedics, Inc.

Depuy P.F.C. E Knee System, stabilized plus tibial insert; polyethylene tibial insert, Sz 2, 20 mm; Ref 96-2714.

DateAug 12, 2005 (initiated May 26, 2005)
StatusTerminated
ReasonAn incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root causeOther
Quantity35
DistributionNationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.