ClearanceBrief
Recall Z-1205-05 ยท NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing

Depuy Orthopaedics, Inc.

Depuy P.F.C. E knee system, TC3 tibial insert; polyethylene tibial insert, Sz 2, 30.0 mm; Ref 96-0517.

DateAug 12, 2005 (initiated May 26, 2005)
StatusTerminated
ReasonAn incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root causeOther
Quantity24
DistributionNationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
FDA recordOpen on fda.gov