ClearanceBrief
Recall Z-1161-05 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Depuy Orthopaedics, Inc.

Depuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 3, 17.5 mm; Ref 1581-13-017.

DateAug 12, 2005 (initiated May 26, 2005)
StatusTerminated
ReasonAn incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root causeOther
Quantity6
DistributionNationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033272DEPUY SIGMA TIBIAL INSERTS
DePuy Orthopaedics, Inc., Feb 5, 2004