ClearanceBrief
Recall Z-1025-05 ยท NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing

Depuy Orthopaedics, Inc.

Depuy brand LCS Complete metal backed patella, porocoat, STD+; REF 1294-09-650.

DateJul 21, 2005 (initiated Jun 7, 2005)
StatusTerminated
ReasonThere may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.
Root causeOther
ActionHospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.
DistributionNationwide and international.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.