Recall Z-1553-05 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump, catalog #2M8064, 2M8064R (refurbished), 2M8064G, 2M8064U, 2M8064F, 2M8064A, 2M8064D & 2M8064B; Baxter Healthcare Corporation, Deerfield, IL 60015

DateSep 21, 2005 (initiated May 3, 2005)
StatusTerminated
ReasonThe mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.
Root causeOther
ActionSafety Alert letters dated 5/3/05 were sent to all Flo-Gard 6201 and 6301 Volumetric Infusion Pump customers. The letters informed the users of the potential for the pole clamp mounting screws to come loose, and recommended that they inspect the Flo-Gard pole clamp during their routine maintenance or as part of the recommended annual preventative cycle. If the pole clamp appears to be loose, the users were provided with the following steps to be followed to ensure that the pole clamp is properly secured: a) The recessed area in the housing for the clamp body must be clean. b) The three moun
Quantity61,759 units
DistributionNationwide including Puerto Rico, and internationally to Canada, Taiwan, Thailand, Belgium, Switzerland, New Zealand, China, Argentina, Peru, Australia, Asia, Hong Kong and the Philippines.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915522FLO-GARD(R) 6201