ClearanceBrief
Recall Z-0939-05 ยท JQP Calculator/Data Processing Module, For Clinical Use

Misys Healthcare Systems

Mysis Laboratory version 6.1

DateJun 28, 2005 (initiated May 10, 2005)
StatusTerminated
ReasonWhen a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more suceptibilities on the culture will be deleted.
Root causeOther
ActionFirm sent Product Safety Notice PSN 05-02 on May 10, 2005 which also provides a procedural work-around and an invitation to obtain a service pack to correct the software flaw.
Quantity21
DistributionNationwide and United Kingdom and Denmark
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.