Recall Z-0915-05 ยท JPA System, Multipurpose For In Vitro Coagulation Studies

Instrumentation Laboratory Co.

ACL Futura, ACL Advance and ACL TOP Coagulation Analyzers used with the HemosIL RecombiPlasTin reagent

DateJun 21, 2005 (initiated May 19, 2005)
StatusTerminated
ReasonPatient prothrombin time (PT) may report low for individuals on anticoaglant therapy
Root causeOther
ActionInstrumentation Laboratory issued an 'Urgent Product Notification' on 5/19/2005 to Beckman Coulter, Miami FL (DIstributor) to provide the written notification and advise customers to review PT curves an any patient on anticagulant treatment who reports a low result , and manually view the curve where a PT is flagged. A software upgrade will be availabel in July 2005.
Quantity617 units
DistributionNationwide Canada