Recall Z-1446-05 ยท HTZ Instrument, Cutting, Orthopedic

Biomet, Inc.

Biomet Vanguard microplasty slidex femoral 4-in-1 block, 60 mm, stainless steel; Ref. 32-485102.

DateAug 31, 2005 (initiated May 16, 2005)
StatusTerminated
ReasonThe instrument may have been assembled improperly, which may result in reversed resection cuts on the femur and affect implant performance, resulting in the need for revision surgery.
Root causeOther
ActionBiomet distributors were issued a letter dated 5/16/05 instructing them to locate the instruments and conduct a field inspection of the instruments to verify the pin location offset and ensure the correct orientation of the anterior marking on the slide component.
Quantity462
DistributionNationwide, Australia, Austria, Belgium, Canada, Costa Rica, Denmark, Finland, Greece, Israel, Japan, Mexico, Netherlands, New Zealand, South Korea, Spain, and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.