Recall Z-0920-05 ยท FNL Bed, Ac-Powered Adjustable Hospital
Carroll Healthcare Inc.
Carroll Healthcare brand Spirit bed; models BH2180-111AE, BH2180-111PE and BH2180-111FE.
| Date | Jun 22, 2005 (initiated May 17, 2005) |
| Status | Terminated |
| Reason | The diodes in the control panel may fail, which could result in the thermal breakdown of the circuit board inside of the control box, and cause the bed to cease functioning. |
| Root cause | Other |
| Action | Consignees were sent a fax on 5/17/05 advising them of the problem and instructing them to lock-out the HILO pendant function on the attendant control pad until the firm visits their facility and corrects their beds. |
| Distribution | Nationwide. |
| FDA record | Open on fda.gov |