ClearanceBrief
Recall Z-0919-05 ยท FNL Bed, Ac-Powered Adjustable Hospital

Carroll Healthcare Inc.

Carroll Healthcare brand Arro bed models BA3180-151 AE, BA3180-151 PE, BA3180-151 FE, BA3180-171 AE, BA3180-171 FE and BA3180-171 FE.

DateJun 22, 2005 (initiated May 17, 2005)
StatusTerminated
ReasonThe diodes in the control panel may fail, which could result in the thermal breakdown of the circuit board inside of the control box, and cause the bed to cease functioning.
Root causeOther
ActionConsignees were sent a fax on 5/17/05 advising them of the problem and instructing them to lock-out the HILO pendant function on the attendant control pad until the firm visits their facility and corrects their beds.
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.