Recall Z-0865-05 ยท DPW Flowmeter, Blood, Cardiovascular

Philips Ultrasound, Inc.

Philips HDI 4000 Ultrasound System Part numbers 9896 053 15031 (8500-0066-01), 9896 053 18811 (8500-0080-01), 9896 053 22011 (8500-0087-01)

DateJun 3, 2005 (initiated May 6, 2005)
StatusTerminated
ReasonThe acoustic display of Mechanical Index and Thermal Index is inaccurate - two transducers may exceed limits for acoustic intensity and MI.
Root causeOther
ActionOn May 6, 2005 the firm mailed a letter, via return receipt requested, to customers. The letter advised of the problem, and provided instructions to create new low power presets. The letter also advised that software would be issued on/before November 2005.
Quantity1225 units
DistributionThe units were distributed worldwide. Domestic distribution included hospitals and medical centers.