Recall Z-0902-05 ยท ITI Wheelchair, Powered

Invacare Corporation

Pronto Series Power Wheelchair, model M51.

DateJun 17, 2005 (initiated May 2, 2005)
StatusTerminated
ReasonIf the arm of the wheelchair is overloaded, the joystick mounting brackett and/or arm pivot may fail due to variability in the plastic used by the supplier to make the components. This failure may cause the user or a caregiver to fall if they are leaning on the arm, and result in injury.
Root causeOther
ActionThe recalling firm contacted all consignees by phone between May 9 and May 25, 2005 and sent an urgent field correction needed letter, dated April, 2005 to all consignees.
Quantity1418 devices.
DistributionThe product was sent to consignees located in AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. There was one foreign consignee in France: Poirer, Zac Papillon, 3720 Parcay Meslay, Corps, F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021680INVACARE POWERED WHEELCHAIRS, MODELS M50, M51, M71 & M91
Invacare Corp., Jun 18, 2002