ClearanceBrief
Recall Z-0241-06 ยท NDH Nail, Fixation, Bone, Metallic

Stryker Howmedica Osteonics Corp.

Gamma3 Long Nail Kit, 120 degree, 11 x 320mm Right. And 125 degree Nail.

DateDec 1, 2005 (initiated Nov 22, 2004)
StatusTerminated
ReasonThe Gamma Nail may have been manufactured with an incorrect lag screw hole angulation. Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees. Catalog number 3220-0320S, lot code K949408 was manufactured with a lag screw hole angulation of 125 degrees instead of 120 degrees.
Root causeOther
ActionNotification letters and Product Accountability forms were sent via FedEx on 11/22/2004 with return receipt.
Quantity19
DistributionThe product was sent to Stryker Orthopaedics branches and agents throughout the US. There was one hospital that was a direct account.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K034002GAMMA3 NAIL SYSTEM
Howmedica Osteonics Corp., Jan 15, 2004