ClearanceBrief
Recall Z-0826-05 ยท DPS Electrocardiograph

Smiths Medical PM, Inc.

Advisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925474325.

DateMay 12, 2005 (initiated Mar 16, 2005)
StatusTerminated
ReasonAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.
Root causeOther
ActionOn March 16, 2005, customers were emailed Safety Action Bulletin 05-003 informing them of the low CO2 (End Tidal and Inspired) measurements in percent (%) and urging them to use mmHg or kPa as the unit of measurement . In addition, a Technical Worksheet will be distributed to customers who request a software upgrade. The Technical Worksheet will instruct them how to perform the software upgrade which corrects the CO2 measurements (End Tidal and Inspired) in percent (%).
Quantity4
DistributionCA and NY within US. Mexico, Argentina, Columbia, Chile, Brazil, Ecuador, Guatemala and Pakistan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010770BCI ADVISOR VITAL SIGNS MONITOR (9200)
Bci, Inc., Jun 12, 2001
K011177BCI ADVISOR VITAL SIGNS MONITOR (9200)
Bci, Inc., May 23, 2001
K030407ADVISOR, MODEL 9200
Bci, Inc., Aug 19, 2003
K982279VITAL SIGNS MONITOR; ADVISOR MODEL # BCI 9200
Bci Intl., Inc., Sep 25, 1998