Recall Z-0848-05 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Medical Systems

Philips Medical HeartStart MRx Monitor/Defibrillator Model: M3535A, M3536A

DateMay 17, 2005 (initiated Apr 29, 2005)
StatusTerminated
ReasonThe device may disarm and not deliver shock when performing a Synchronized Cardioversion
Root causeSoftware design
ActionPhilips Medical Systems notified accounts by letter on 5/2/05. The firm provided instructions for identifying the effected devices and actions to be taken until the software is availablel for installation.
Quantity5125 units
DistributionNationwide Foreign: Canada, Austria, Australia Bahrain, Belgium, Brazil, China, Hong Kong, India, Finland,France, Germany, Hungry, Israel, Italy, Korea, Kuwait, Latvia, Lebanon, Mexico, Namibia, Libya, Netherland, New Zealnd, Norway, Oma, Philippines, Portugal, Qatat, Russia, South Africa, Singapo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003