Recall Z-0866-05 ยท LEG Enzyme Immunoassay, Valproic Acid

Ortho-Clinical Diagnostics

VITROS VALP Reagent , REF 680 1710. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Responsible firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, N.Y. 14626.

DateJun 4, 2005 (initiated Apr 18, 2005)
StatusTerminated
ReasonNegatively biased results may be observed when the reagent is stored on the VITROS 5,1FS Chemistry System for less than seven days.
Root causeOther
ActionLetters dated 4/18/2005 with instructions to process a mid-level QC fluid every eight hours when processing patient samples for valproic acid. Replacement product to be shipped as soon as available.
Quantity139 boxes
Distribution37 U.S. consignees nationwide. Ortho-Clinical foreign affiliates in England, France and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042476VITROS CHEMISTRY PRODUCTS VALP REAGENT; VITRO CHEMISTRY PRODUCTS CALIBRATOR KIT 12; VITROS CHEMISTRY PRODUCTS TDM PERFOS
Ortho-Clinical Diagnostics, Inc., Dec 1, 2004