Recall Z-0921-05 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems

Flexible Implant tube, blind end, 350 mm, part # GM11003730, labeled in part ***Varian Medical Systems, Inc 3100 Hansen Way Palo Alto, CA 94304 USA www.varian.com***. Available as single unit or contained in the Flexible Interstitial Implant Tube Set, part number GM11004680. For use with the Varian GammaMed Plus brachytherapy afterloader device.

DateJun 22, 2005 (initiated Apr 14, 2005)
StatusTerminated
ReasonA defective treatment applicator in a cancer therapy medical device may become separated potentially causing patient injury.
Root causeOther
ActionThe recalling firm notified end users by Customer Technical Bulletin on 4/14/05. The bulletin advised of potential for the blind end of the implant tube to become separated with the possible effect of the tube end being inhaled by a patient. Consignees were instructed to stop using product and return for replacement. The firm''s bulletin is also available on the company website.
Quantity33 devices in domestic distribution
DistributionProduct was distributed to direct accounts in CT, MI, PA, CO, VT and MO. Product was additionally sold to foreign accounts in internationally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983436GAMMAMED PLUS HIGH DOSE RATE REMOTE AFTERLOADING SYSTEM
Gammamed USA, Oct 29, 1998