ClearanceBrief
Recall Z-1596-05 ยท NBW System, Test, Blood Glucose, Over The Counter

Lifescan Inc

LifeScan brand OneTouch Ultra Meter - Owner''s Booklet; Product is manufactured and distributed by LifeScan, Inc., 1000 Gibraltar Dr., Milpitas, CA 95037

DateSep 29, 2005 (initiated Nov 21, 2003)
StatusTerminated
ReasonLifeScan brand OneTouch Ultra Meter - Owner's Booklets are missing instructions on how to set the Units of Measurement (UOM)--mg/dL versus mmol/L. This error has the potential to lead to acute hyperglycemic complications if the user fails to realize that 18mg/dL of glucose is equal to 1 (one) mmol/L.
Root causeOther
ActionFirm contacted all accounts by letters of 11/21/03, 2/4/2004, 2/17/04 and 2/9-10/2004 providing important information regarding the owner''s booklet.
QuantityApproximately 572,000 units
DistributionThe recall product was distributed in the U.S. only. LifeScan manufactured and distributed OneTouch Ultra System Kits with the affected OneTouch Ultra Meter Owner''s Booklets between March 31, 2003 and October 3, 2003.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K024194ONE TOUCH ULTRA AND ONE TOUCH INDUO BLOOD GLUCOSE MONITORING SYSTEMS
Lifescan, Inc., Feb 7, 2003