Recall Z-0943-05 ยท FMG Stopcock, I.V. Set

Baxter Healthcare Corp.

Baxter Interlink System Continu-Flo Solution Set with Large Bore 4-Way Stopcock Extension Set, 10 drops/mL, product code 3C0062s; a sterile, nonpyrogenic fluid pathway for administration of IV fluids; Manufactured by an affiliate of Baxter Healthcare Corporation, Deerfield, IL 60015 USA, Made in Costa Rica

DateJun 30, 2005 (initiated Apr 21, 2005)
StatusTerminated
ReasonThere is an unsecure connection between the male Luer and the stopcock on the I.V. extension set.
Root causeOther
ActionRecall letters dated 4/21/05 were sent to the direct accounts via first class mail on the same date to the attention of the Director of Materials Management. The accounts were informed of the reports of an unsecure connection between the male luer and the stopcock due to silicone on the male luer tip, and were requested to examine their inventory for the affected lots of product, and call Baxter Healthcare Center for Service at 1-888-229-001 to return the affected product for credit.
Quantity107,832 sets
DistributionNationwide and internationally to American Samoa and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022895ELCAM STOPCOCKS AND MANIFOLDS
Medical Device Consultants, Inc., Oct 18, 2002