Recall Z-0846-05 ยท LNH System, Nuclear Magnetic Resonance Imaging

Hitachi Medical Systems America Inc

Altaire, Magnetic Resonance Imaging Device.

DateMay 17, 2005 (initiated Nov 18, 2004)
StatusTerminated
ReasonThe absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor.
Root causeOther
ActionThe firm implemented site inspections of customers beginning on 11/19/04 which have been completed. Customers that had the suspect component had it replaced.
Quantity189 devices
DistributionDirect accounts located in AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002420HITACHI ALTAIRE
Hitachi Medical Systems America, Inc., Oct 17, 2000
K022390HITACHI ALTAIRE
Hitachi Medical Systems America, Inc., Oct 11, 2002