Recall Z-0846-05 ยท LNH System, Nuclear Magnetic Resonance Imaging
Hitachi Medical Systems America Inc
Altaire, Magnetic Resonance Imaging Device.
| Date | May 17, 2005 (initiated Nov 18, 2004) |
| Status | Terminated |
| Reason | The absorber unit in the helium compressor which is located in equipment rooms, and is not accessed by patients or system operators, may fail under pressure, causing an explosion that damages the compressor. |
| Root cause | Other |
| Action | The firm implemented site inspections of customers beginning on 11/19/04 which have been completed. Customers that had the suspect component had it replaced. |
| Quantity | 189 devices |
| Distribution | Direct accounts located in AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI. |