Recall Z-0802-05 ยท LWS Implantable Cardioverter Defibrillator (Non-Crt)

ela Medical Llc

Sentinel Implantable Carioverter-DefibrilIator (ICD) model 2012 is a multi-programmable, two-zone, tiered-therapy device that incorporates anti-tachycardia pacing (ATP), cardioversion, and defibrillation (DEF) tachyarrhythmia therapies. The device also includes a programmable Hot Can electrode and bradycardia pacing support.

DateMay 6, 2005 (initiated Jan 25, 2005)
StatusTerminated
ReasonA change in the automatic capacitor reform schedule during the later portion of the device's useful life can result in excessive 750V charge times and disable the device's ability to report an End of Life (EOL) indicator due to excessive charge times unless the patient is receiving periodic 750V therapy shocks.
Root causeOther
ActionAngeion notified physicians responsible for managing each remaining implanted ICD, also provided them with programming means to change the automatic cap charging schedule, and to monitor and document the reprogramming process.
Quantity3 in US and 2 OUS
DistributionAZ, CA, GA, IL, IN, KY, MA, MD, MI, MN, NC, NJ, OH, PA and to UK, Italy, Germany and Argentina.