Recall Z-0842-05 ยท DXE Catheter, Embolectomy

Arrow International Inc

The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistula

DateMay 13, 2005 (initiated Mar 18, 2005)
StatusTerminated
ReasonThe orange Pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.
Root causeOther
ActionThe firm began notification of consignees on 3/23/2005 via certified letter.
Quantity2418 UNITS
DistributionProduct was distributed nationwide to dealers, hospitals, sales representatives, and subsidiaries (distributors). The product was also distributed to Australia, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Japan, Netherlands, Russia, South Africa, South Korea, Spain, Sweden, Ta
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011056ARROW-TREROTOLA PERCUTANEOUS THROMBOLYTIC DEVICE OR PTD
Arrow Intl., Inc., Oct 24, 2001