Recall Z-0928-05 ยท BSP Needle, Conduction, Anesthetic (W/Wo Introducer)

Becton Dickinson & Company

BD 25G Spinal Needles. (Spinal Anesthesia Needle)

DateJun 25, 2005 (initiated Mar 16, 2005)
StatusTerminated
ReasonA significant level of 25G spinal needle hub leakage complaints was observed. Subsequent investigation found the leakage to be attributable to improper formation of the spinal needle hub, specifically an undersized taper and the prescence of a slight sink in the taper surface.
Root causeOther
ActionThe firm sent out notification letters on 3/16/2005.
Quantity801580 units
DistributionThe firm uses distributors and ships the product directly to over 4000 hospitals and clinics nationwide.